Weschler Instruments – Manufacturer & distributor of measurement & control equipment
440.238.2550   /   800.557.0064   /   
analog meters, digital meters, transducers, signal conditioners, data loggers and test instrumentation
Home | About Us | What's New | Industries | Weschler Products | All Products | Meter Mod Center | Reference | Contact | Site
Search Catalog 
      GO   
 All Categories    Multi-Function Calibrator - Martel
Weschler Instruments is a manufacturer and distributor of electrical and electronic measurement equipment. Customers worldwide use our products in power plants, steel mills, chemical plants, paper mills and other industrial applications that require rugged, reliable and accurate instrumentation.
All Categories > Test Equipment > Martel Beta Process Calibrators > Multi-Function Calibrator - Martel > Item # MA/MC-1200  
    Download PDF Download PDF     Printable Page     Email Email This Page     Save To Favorites Save To Favorites
Request Information

Item # MA/MC-1200, Multi-Function Calibrator with test leads, 4 AA batteries, NIST certificate



Multi-Function Calibrator

larger image
  • Measure and source T/Cs, RTDs, Ohms, current, voltage, frequency; source pulse trains
  • Isolated mA/V read-back for complete transmitter calibration
  • Pressure module communication port
  • Built-in 24 V supply drives 4–20mA loops
  • Direct entry of custom RTD coefficients
  • All source modes can be programmed with setpoints to speed linearity tests
  • Highest accuracy in class (to 0.015% rdg.)
A variety of temperature probes are available.





Specifications

Read and Source

Read Isolated: 0.000 to 30.000 VDC (±0.015% of rdg ±2 µV acc), Current: 0.000 to 24.000 mA (±0.015% of rdg ±2 µA acc.)
Read Non-isolated: 0.000 to 20.000 VDC (±0.015% of rdg ±2 µV acc), Current: 0.000 to 24.000 mA (±0.015% of rdg ±2 µA acc.)
Source, Voltage: 0.000 to 20.000 VDC (±0.015% of rdg ±2 µV acc), Current: 0.000 to 24.000 mA (±0.015% of rdg ±2 µA acc.)

Frequency (1-20 V amplitude)

CPM Source & Read: 2.0 to 600.0 CPM
Hz Source & Read: 1.0 to 1000.0 Hz
kHz Source & Read: 1.0 to 10.00 kHz

Pulse Source (1-20 V amplitude)

Pulses: 1 to 30,000.0
Rate: 2 CPM to 10 kHz

Read and Source

J: -200.0 to +1200.0 °C, Accuracy: 0.2
K: -200.0 to +1370.0 °C, Accuracy: 0.3
T: -200.0 to +400.0 °C, Accuracy: 0.2
E: -200.0 to +950.0 °C, Accuracy: 0.2
R: -20.0 to +1750.0 °C, Accuracy: 1.2
S: -20.0 to +1750.0 °C, Accuracy: 1.2
B: +600.0 to +1800.0 °C, Accuracy: 1.2
C: 0 to +2316.0 °C, Accuracy: 0.6
XK: -200.0 to +800.0 °C, Accuracy: 0.2
BP: 0 to +2500.0 °C, Accuracy: 0.9
L: -200.0 to +900.0 °C, Accuracy: 0.2
U: -200.0 to +400.0 °C, Accuracy: 0.25
N: -200.0 to +1300.0 °C, Accuracy: 0.4
Ni120 (672): -80.0 to +260.0 °C, Accuracy: 0.2
Pt100 (385): -200.0 to +800.0 °C, Accuracy: 0.2
Pt100 (3926): -200.0 to +630.0 °C, Accuracy: 0.2
Pt100 (3916): -200.0 to +630.0 °C, Accuracy: 0.2
Pt200 (385): -200.0 to +630.0 °C, Accuracy: 0.8
Pt500 (385): -200.0 to +630.0 °C, Accuracy: 0.4
Pt1000 (385): -200.0 to +630.0 °C, Accuracy: 0.2
Cu10: -100.0 to +260.0 °C Accuracy: 1.4
YSI400: +15.00 to +50.00 °C Accuracy: 0.1
Cu50: -180 to +200 °C Accuracy: 0.4
Cu100: -180 to +200 °C Accuracy: 0.3
Pt385-10: -200 to +800.0 °C Accuracy: 1.4
Pt385-50: -200 to +800.0 °C Accuracy: 0.4

Thermocouple mV

Read & Source: -10.000 to +75.000 mV, Accuracy: ±0.02% of rdg, ±10 µV

Ohms

Read: 0.00 to 4000.0ohm
Source: 5.0 to 4000ohm

Operating Temperature

-10 to 50 ºC

Stability

±0.005% of reading/ °C outside 18-28 °C

Dimensions

8.7 x 4.2 x 2.3 in


 DOWNLOADS 

 width= Martel Beta Process Calibrators Catalog Page
(PDF, 179KB)


 ACCESSORIES 


null

Carrying Case

Request Information