Weschler Instruments – Manufacturer & distributor of measurement & control equipment
440.238.2550   /   800.557.0064   /   
analog meters, digital meters, transducers, signal conditioners, data loggers and test instrumentation
Home | About Us | What's New | Industries | Weschler Products | All Products | Meter Mod Center | Reference | Contact | Site
Search Catalog 
      GO   
 All Categories    Stylist - Surface Mount Taut Band 260 - AC Voltmeter (Iron Vane) Configurator - Yokogawa
Weschler Instruments is a manufacturer and distributor of electrical and electronic measurement equipment. Customers worldwide use our products in power plants, steel mills, chemical plants, paper mills and other industrial applications that require rugged, reliable and accurate instrumentation.
All Categories > Product by Manufacturer > Yokogawa > Yokogawa AC Current and Voltage Meters > Stylist AC Voltmeters - Yokogawa > Stylist - Surface Mount Taut Band 260 - AC Voltmeter (Iron Vane) Configurator - Yokogawa 
    Download PDF Download PDF     Printable Page     Email Email This Page     Save To Favorites Save To Favorites

Stylist - Surface Mount Taut Band 260 - AC Voltmeter (Iron Vane) Configurator - Yokogawa


Configure ProductRequest Information


ac-blue-stylus

larger image
AC Voltmeter
Stylist Surface Mount Panel Meter, Model 260.
Pivot & Jewel - In this design the coil and pointer are supported by polished steel pivots at each end which fit into jewel bearings. Pivot and jewel meters are recommended for high- vibration environments because of the mechanism’s inherent stability and strength. Accuracy and repeatability are almost equal to the taut-band’s, and good performance from the pivot and jewel construction is the reason for its long popularity. Don’t specify pivot and jewel meters for applications that involve high shock, but do take advantage of this meter’s ability to withstand vibration.



 Product Overview  

Accuracy

All DC and AC iron vane, ±2% of full scale, AC rectifier type; ±3% of fullscale (with 60 Hz sine wave at 25°C). Expanded scale voltmeters ±0.5% of full scale.

ANSI Specification

All meters meet ANSI Specifications C-39.1

Overload

Voltmeters (AC + DC) - 50% momentary, 20% sustained. Ammeters (AC+DC) momentary: 10 times the rated current for 10 consecutive intervals of .5 seconds with 1 minute interval between successive applications; sustained: 20% for six hours.

Insulation Level

2600V/1 min. (2-1/2-4-1/2)
1500V/1 min. (1-1/2")

Response Time

3 seconds (Max.) - Microammeters
2 seconds (Max.) - All other DC ratings
2.5 seconds (Max.) - All AC ratings
Maximum 40% overshoot

Burden Data

AC Ammeter 5 amps @60 Hz, 0.5VA, Max. 0.5 Power factor lagging
AC voltmeter - unity power factor